4.2 Measurement in FULL AUTO Mode
4.2 Measurement in FULL AUTO Mode
There are two modes of performing measurement; FULL AUTO and MANUAL.
FULL AUTO mode is selected automatically when the power is turned ON or when the
PRINT switch is pressed.
The examiner can also choose FULL AUTO mode with the AUTO switch.
After displaying the examinee’s eye on the monitor, the examiner presses the START switch.
The instrument will then automatically perform the alignment, focusing and measurement.
It is also possible to have the instrument automatically measure the left eye successively
after measuring the right eye, and then print out the results.
In order to use the fully automatic function, set the options in SET mode ( sections 6.1 and
6.2) as follows:
PRINT: ON
R–L MEASURE: ON
AUTO PRINT: ON
Select MANUAL mode with the MANU. switch when the pupil is eccentric, or when measur-
ing a certain examinee is apt to result in error.
See section 4.3 for the procedure for manual measurement.