CANON Digital Galvano Scanner System GM-1000 Series
Users Manual 1.20
102
Command ID 20
Command
Name
Target Position Setting
Data Target position (Pulse)
Return
Value
0: Command execution successful
1: Command execution unsuccessful
Explanation This command is used to set a target position by step movement.
Related
Command
Step movement procedure
Command ID = 10: Target Value Setting Mode
Command ID = 20: Target Position Setting
Command ID = 8 Data = 0: Step Movement Start
Command ID 22
Command
Name
Target Velocity Setting
Data Target velocity (Unit: pulse/s)
Return
Value
0: Command execution successful
1: Command execution unsuccessful
Explanation This command is used to set the maximum velocity of step movement.
The velocity cannot be higher than “Parameter ID = 2: Maximum velocity.”
If this command is not set, the velocity specified by “Parameter ID = 2: Maximum
velocity” is used.
Related
Command
Step movement procedure
Command ID = 22 Target Velocity Setting
Command ID = 10: Target Value Setting Mode
Command ID = 20: Target Position Setting
Command ID = 8 Data = 0: Movement Start