4.2.1. Step Movement Responce Time Measurement
Control Software has a function that measures the step movement response time. This function is can
only be executed by the control software. It cannot be executed by a RS-232C command.
Put a check in the Enable Scope checkbox
Select the time range for step movement response to be measured
(As the response time depends on the displacement, change when necessary.)
Specify the displacement (angle and pulse count) for relative position (STEP) and absolute position.
Press the STEP button (either positive or negative position direction can be specified) or GO button.
The Scope and Measurement window appears automatically